10.
Attaching the Ameda Freezer Bag
Using the HygieniKit, you may pump directly into Ameda Freezer Bags (sold seperately).
|
Using the HygieniKit
|

- Insert the neck of the breast flange into the top of the freezer bag.
|

- Pleat and fold the excess bag material arount the neck of the breast flange.
|

- Place the sealing tie into the grove used for the collar of the handle assembly and twist tightly.
|
|
Top 
11.
The Flexishield Areola Stimulator
The Flexishield Areola Stimulator is an optional accessory that can be purchased separately and used with the HygieniKit to provide increased stimulation and encourage the
let-down reflex. The Flexishield is designed to compress the areola area, helping to stimulate a let-down reflex. It is especially useful for smaller-breasted women who may benefit
from added stimulation.
|
- Insert the Flexishield into the breast flange. The rings on the Flexishield should contact the nipple tunnel walls once it is fully inserted.
|
|
- Seal the Flexishield to the breast flange by locking the edge of the Flexishield over the rim of the breast flange, using the palm of your hand
or thumb and forefinger.
|
Top 
12.
Technical Data
Power Requirements
AC Adapter and Car Adapter (included)
AC Adapter - North American Plug Configuration
9VDC/1.3A, 120V/60Hz, Hollister Stock No. 622401, direct plug in AC power adapter.
AC Adapter - Australian Plug Configuration
9VDC/1.3A, 120V/60Hz, Hollister Stock No. 402707, direct plug in AC power adapter.
Car Adapter (included in backpack style only)
Otherwise sold seperately as Hollister Stock No. 17079).
12VDC, Hollister Stock No 17079, 15A/250V fused adapter.
Batteries (not included)
Six AA batteries, 1.5V alkaline or 1.2V rechargeable.
Regulatory Compliance
CE conformity to Medical Device Directive 93/42/EEC.
EN 60601-1 1996 Medical Electrical Equipment
EN 60601-1-2 1994 EMC
UL and CSA approved with North American power adapter.
UL 1431 Personal Hygiene and Health Care Appliances
File Number E189700
CSA Std. C22.2 No 125-M1984 Electromedical Equipment
Class 8711 01 Electromedical Equipment - Risk Class 2
Class 3872 01 Hygiene and Cosmetic Appliances
File Number LR76525I
Top 
13.
Troubleshooting the Purely Yours Breast Pump
Interruption of Electrical Service
If you experience an interruption in the normal operation of the electrical and/or mechanical functioning of the pump, simply do the following:
- Unplug the AC adapter cord from the electrical outlet.
- Wait a few seconds, then plug the AC adapter cord back into the electrical outlet.
Decreased Suction
If it feels like the suction has decreased, try the following:
- Remove the tubing and white connector from the side of the pump and place a finger over the pump outlet. If you feel suction, the pump is operating properly, but the collection kit may
not be correctly assembled or the white valve may be damaged. If no suction is felt on your finger, try unplugging the adapter from the electrical outlet, wait a few seconds, and plug it
back into the electrical outlet.
- Make sure the collection kit has been properly assembled. Check the assembly information in the section
Using the HygieniKit.
- Inspect the white valve for a tear or pinhole. This component is critical to achieving proper suction. Replace the white valve with one of the spare valves provided with the product.
- Check the tubing adapter on the side of the pump to make sure it is closed when single pumping. If open, close the adapter.
|